GL Optic Sp. z o.o. / Integrating spheres
GL OPTI SPHERE 1100
GL OPTI SPHERE 1100
For luminous flux measurement and radiant power of mid-size LED luminaires and large LED modules we designed an integrating sphere system
Price on request
Available to order
✓Expert selection supportWe help verify suitability before purchase.
✓Calibration optionsConfigure calibration with the instrument.
✓Instrument lifecycleManage serial number, history and certificates.
EXPERT VIEW
Why this instrument?
For luminous flux measurement and radiant power of mid-size LED luminaires and large LED modules we designed an integrating sphere system All published technical parameters below are transcribed from the current official manufacturer page. Final system configuration is verified with the quotation.
PRODUCT FAMILY
GL OPTI SPHERE variants
Current professional GL Optic Integrating spheres product family.
TECHNICAL DATA
Specifications
Technical data
Spectral range*340 – 1700 nm
Sphere inner diameter1100 mm
Entrance aperture diameter168 mm
Sphere materialComposite
Inner coatingBarium Sulfate (BaSO4) high-reflectance material (R98)
Outer coatingBlack finish
Reflectance properties97%
Auxiliary light sourceWhite LED
Spectrometer portDirect connection or SMA fiber-optic
Standards complianceCE, LM 79, CIE 127:2007 CIE S 025/E:2015
Maximum DUT dimensions in accordance with CIE S 025/E:2015100 mm (diameter or diagonal)
Maximum dimension for optimal measurement (1/3 x sphere diameter)330 mm (diameter or diagonal)
Maximum DUT weight3 kg
Sphere frameHinged
Sphere center positioningN/A
Mechanical breadboard with postFor 4π measurement
USB source controller for auxiliary light sourceWith current source and relay switch for external power supply
Universal post with standard lamp socketsUniversal DUT fixing table (breadboard) for measurement in 4π geometry
External dimensions [W x H x D]1260 mm x 1800 mm x 1220mm
Weight60 kg
GL OPTI SPHERE 1100Request quote