Professional Light & Color MeasurementInstruments · Calibration · Expert support
Catalog previewThis instrument is visible for evaluation and quotation, but publication review is not yet complete.
GL Optic Sp. z o.o. / Goniophotometers

GL GONIO SPECTROMETER GLG 30-1800

GL GONIO SPECTROMETER GLG 30-1800

Automated laboratory level gonio spectrometer generates spectral and color data at any angle, accurately characterize big and small fixtures.

Price on request
Available to order
✓Expert selection supportWe help verify suitability before purchase.
✓Calibration optionsConfigure calibration with the instrument.
✓Instrument lifecycleManage serial number, history and certificates.
Quick quotation
EXPERT VIEW

Why this instrument?

Automated laboratory level gonio spectrometer generates spectral and color data at any angle, accurately characterize big and small fixtures. All published technical parameters below are transcribed from the current official manufacturer page. Final system configuration is verified with the quotation.

Professional useTechnical supportCalibration ready
PRODUCT FAMILY

GL GONIO variants

Current professional GL Optic Goniophotometers product family.

TECHNICAL DATA

Specifications

Technical data

CIE Goniometer typeFar Field Type C with horizontal optical axis
Angular range C axis± 180°
Angular range γ axis± 180° theoretical (usable depending on geometry due to shadowing)
Angular positioning precision0.1°
Reproducibility C axis0.004°
Reproducibility γ axis0.004°
Angular speed C axisup to 45°/s
Angular speed γ axisup to 45°/s
DUT photometric center positioning700 mm, stepper motor operated
Photometric distance9 m – 27 m Distance >= largest DUT dimension x (5 to 15) (CIE S 025/E:2015)
DUT mounting plateRound d=250 mm
(bread board)Fixing: multiple mounting holes: M3, M4, M5, M6; chuck with sliding blocks
Maximum DUT dimension≤ 1800 mm (diameter or diagonal)
Goniometer dimensions800 x 1714 x 968 mm (W x H x D)
Maximum operating footprint diameter2400 mm
Maximum load30 kg
Goniometer weight360 kg
Power supply and max. consumptionPC connection by Ethernet; 110-230 V 600 W
Sensor typeGL OPTIC Spectrometer range:
GL GONIO SPECTROMETER GLG 30-1800Request quote
Request quote